Thermal variations in the patient compartment of an emergency ambulance: A feasibility study in an Irish context.

Damien Gaumont, Claire Armstrong, Gordon Armstrong

DOI: http://dx.doi.org/10.32378/ijp.v3i2.147

Abstract

Background

The unpredictability of the nature of the next call is a basic feature of Emergency Services; the call could vary from a trauma victim, to a hypothermic patient or a prehospital birth. All patients (other than those who are pyrexic) have in common the need for a warm environment to prevent deterioration in their condition. Multiple observation studies found that patients suffering from various levels of trauma, arrived in the Emergency Department with hypothermia. Hypothermia, a core temperature <35°C, affects multiple organ systems, and is associated with poor outcomes including death. Also, cold has been reported as negatively impacting the comfort of an ill or injured patient. It is currently assumed that the ambulance patient compartment’s heater (Air Top Evo 40, Webasto™, Gilching, Germany), produces enough heat to offer thermal comfort and to help prevent further decrease of body temperature in the hypothermic patient. However, what is not clear is for how long and to what ambient temperature the ambulance’s patient compartment needs to be heated, to provide the ambulance’s furniture with sufficient stored energy to maintain the patient at an appropriate temperature for the duration of their transport to hospital. We consider how current practices and behaviours may need to be adapted to improve patient comfort and outcomes.

Objectives

This study is to determine the feasibility of measuring and monitoring temperatures in a new generation Emergency Ambulance. The overarching objective, is to optimise patient comfort, outcome and prevention of hypothermia.

Methods

Using thermocouples, a data logger and a thermal camera to record temperatures at strategic locations in the patient compartment, we recorded the variation of temperature in a typical new generation Emergency Ambulance compliant with the CEN - EN 1789:2007 standard. Thermal imaging and temperature logging studies were conducted on in May/July 2018. Temperature was logged for 24 hours. The locations examined were the stretcher mattress surface, low and high blanket storage lockers and the outdoor ambient air. The vehicle was located outdoor, facing west-north-west.

Anticipated outcome

This study will provide us with data that can be used to improve patients’ thermal comfort through behaviour and practice change.

Results

The thermal camera images show a significant variation of surface temperature throughout the patient compartment. Preliminary temperature logging experiments show a measurable difference in temperatures at the areas of interest relative to the outside temperature over 24 hours. Some variations in rates of cooling and warming in each area have been observed during the cool – heat period; the stretcher mattress is the slowest to rewarm.

Conclusion

The proposed method of measuring temperature variation in targeted locations in the patient compartment of a new generation ambulance proves efficient and could be used in further studies.


Keywords

thermal comfort, ambulance design, hypothermia

Full Text:

PDF (poster) PDF

References

Aléx, J. Karlsson, S. and Saveman, B-I. (2013) ‘Patient’s experience of cold exposure during ambulance care’, Scandinavian Journal, Resuscitation and Emergency Medicine, 21(44), available: doi:10.1186/1757-7241-21-44

Haverkamp, F., Giesbrecht, G., Tan, E. (2018) ‘The prehospital management of hypothermia- an up-to-date overview’, Injury, 49,149-164, available: doi: https://doi.org/10.1016/j.injury.2017.11.001

Lapostolle, F., Sebbah, J.L., Couvreur, J., Koch, F.X., Savary, D., Tazarourte, K., Egman, G., Mzabi, L., Galinski, M. and Adnet, F. (2012) ‘Risk factors for onset of hypothermia in trauma victims: the HypoTraum study’, Critical Care, 16(4), R142, available: doi: 10.1186/cc11449

Rogalsky, A. and Chrzanowski, K. (2002) ‘Infrared devices and techniques’, Opto-electrinics review, 10(2), 111-136, available: http://antonirogalski.com/wp-content/uploads/2011/12/Infrared-devices-and-techniques.pdf [accessed 12 June 2018].

Watts, D.D., Roche, M., Tricario, R., Poole, F., Brown, J.J., Colson, G.B., Trask, A.L. and Fakhry, S.M. (1999) ‘The utility of traditional prehospital interventions in maintaining thermostatis’, Prehospital Emergency Care, 3(2), 115-122 available: doi: https://doi.org/10.1080/10903129908958918.

Zafren, K. et al (2014) ‘Wilderness Medical Society Practice Guidelines for the Out-of-Hospital Evaluation and Treatment of Accidental Hypothermia: 2014 Update’, Wilderness and Environmental Medicine, 25(4) Supplement, S66—S85, available: doi: https://doi.org/10.1016/j.wem.2014.10.010.



DOI: http://dx.doi.org/10.32378/ijp.v3i2.147

Refbacks

  • There are currently no refbacks.




Copyright (c) 2018 Damien Gaumont, Claire Armstrong, Gordon Armstrong

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.


IJP is an Open Access journal and articles published are distributed under the terms of the Creative Commons Attribution-NonCommercial License (CC BY-NC). Article copyright remains with individual authors. Authors grant unrestricted copyright license to the IJP. Readers may copy, distribute, build upon, reproduce, and display the work for non-commercial purposes with the proper citation of the original work in this publication. DPLA Rights statement: http://rightsstatements.org/page/InC-NC/1.0/?language=en. The IJP supports the National Principles for Open Access Policy Statement.

Electronic ISSN: 2009-938X.

We deposit a copy of all Irish healthcare related articles with Lenus, the Irish Health Repository and eDepositIreland, ensuring a full-text copy is always available and securely stored. The IJP is also listed with DOAJGoogle ScholarScienceOpenWorldCat, Index Copernicus, OAJI and JournalTOCs.